FYS eye exercise device
Welcome you to use the FYS 2000-II vision frequency spectrum apparatus (FYSEED)fys

Table of contents

Chapter One - product briefs
1.1 Product briefs
1.2 Treatment principle
 1. Causes of eye disorders
 2. Visual frequency spectrum music therapy
1.3 Product function
1.4 Applicable scope
1.5 Important prompt

Chapter Two - operating instructions
2.1 Main machine and modules
2.2 Installation prompt
2.3 Power on  procedure
2.4 Application method
 1. Method of treatment
 2. Treatment time
 3. Treatment period
 4. Treatment standard
 5. Attentions

Chapter Three - product maintenance
3.1 Product maintenance
3.2 Technical specification
3.3 Trouble Shootings


Chapter One - product briefs

1.1 Product briefs

This device is based on optics according to the human body vision
physiology research newest achievement and the eye the elementary
theory, its a combination of previous several generation of products
characteristics, its a new generation binocular eye exercise device uses the microcomputer technical design.

This instrument uses 640 nanometer human bodies specific visions frequency spectrum
light to massage eyes and the highly effective activate optic cells and
the human brain optic nerves, it can promote eye organ's blood
circulation, to improve vision conduction function,
to enhance the visual fusion frequency, glitter the change using music
melody synchronization control frequency spectrum light, thereforce
to induce the pupillary reflex and the crystalline lens adjustment, and to
improve eye adjustment function, effectively treats the control
character nearsightedness, ambliopia, to eliminate eye strain, and to delay the presbyopia.

This instrument has gathered the,visual receptivity training, the vision resolution training,
the darkroom training, the eye muscle adjustment training, the visual sense of hearing coordination training, the
stereoscopic vision training and the binocular coordination training - seven effective syntheses training methods,
it has Characteristics of treatment principle scientific, excellent curative effect, highly automatic, vivid interesting
, safe and reliable, it does not have the side effect, it meets the the medical instrument
standard of vision health care treatment apparatus of China.

1.2 Treatment principle

 1. Causes of eye disorders
The eye is the visual organ of human body. The eyeball is composed of photosensitive system and refraction focusing
system, its similar to a camera system which ocnstructed by the focusing to be partial and the photosensitive part.

The eyeball transports and relaxes the ligament with the eyelash shape myo-contraction to adjust the thickness of
the crystalline lens, when look at the distance the ciliary body will be relaxed, when look at near object the ciliary
body will be contracted, thus the adjustment focal point and the distance, caused the physical image correctly to fall
on the retina, its function is equal in the camera focusing; Falls on the retina, the physical image through
the visual cell, receives it and transforms the electrical signal to transmit to cerebral cortex regarding
sense center, this is how we see the color rich object clearly, its function is equal in the camera the photosensitive negative.
Visual image formed separately by focusing system and photosensitive system indispensable, if the crystalline lens and the eyelash
shape myo- appears the adjustment barrier, then may happens the nearsightedness or the hyperopia; If retina and cerebral
cortex vision center growth flaw, then may happens Ambliopia; In addition, by barrier and so on nearsightedness,
hyperopia also may initiate the occurrence of the Ambliopia, and in another way, the Ambliopia occurrence may also
cause the adjustment barrier appearance.
Therefore, if you expect good vision, must guarantee holds in the visual system at the same time two parts complete
and is normal.

 2. Visual frequency spectrum music therapy
research results indicated that, the retina organizes to regard in the
awl cell, includes feels red, is green, the blue three kind of basic
colors photosensitive ingredient, specific wave length
With the intensity red, green, the blue visual frequency
spectrum light (wave length 640 nanometer thioindigo red look at each
other the awl cell to carry on in particular) when the rhythm
stimulates, regards the awl cell specially active and is sensitive,
corresponding causing
Regards certainly the nerve and the cerebral cortex visual main
center is highly excited, at the same time, the frequency spectrum
light glitters the change, can the force induce the pupil conditioned
reflex and the crystalline lens adjustment, the improvement eye
adjustment function.
When uses the delightful music synchronization control frequency
spectrum light the rhythm change, as a result of visual and the sense
of hearing synchronization, can further strengthen 交感神经 the
system excitability and the coordination.

The visual frequency spectrum music therapy is a set of science but standard
vision health care and treatment plan developed and designed based on this newest research
result.

It selected the specific visual frequency spectrum light which result the best treatment effect
, determined the appropriate frequency spectrum light intensity, selected and
matched the melody change delightful interesting to listen to music,
moreover has formulated the science treatment procedure. Guarantees
with the traditional vision The healthy method of treatment compares, its biggest difference
is, in the instrument use process, can is organic the eye focusing
system flexible training and the photosensitivity system function
enhance training unifies together, makes two parts of functions to promote
mutually, enhances unceasingly.

1.3 Product function
Treatment: Types of youth Ambliopia, young people adjustable nearsightedness;
Health care: To eliminate eye strain, prevents the
nearsightedness and ambliopia, delays the presbyopia;


1.4 Applicable scope
(1) the young people nearsightedness patient (2) children ambliopia
(3) eye strain patient (4) presbyopia patient

* Attention: Glaucoma, the eyeground hemorrhage, the retina fall off
and pathological change eye disease patient
is forbid to use, or under the instruction of
doctor.

1.5 Important prompt
 1. The modules only may be used by the manufacturer assigned technical specification;
 2. When uses the battery as power source, when the
power indicator light changes from green to yellow, green in turn glitters,
when finally turns red, indicated the battery capacity is
insufficient, should promptly replace newly
battery, suggested to use high energy environmental protection
battery;
 3. When clean windows lens, avoids using hardly, in order to avoid
scratches the mirror surface;
 4. The instrumentation must install the protection blinkers eyeshade(or eyepatch).
The eyepatch to be supposed the specialist
special-purpose and dismantles the clean regularly, in order to avoid
eye infectious disease overlapping infection;
 5. If appears the malfunction, you should find properly method according to this handbook.
If still could not work normally, go to the special maintenance point or contact the seller
for maintaince service. Disassembles this product arbitrarily not in the row of the maintenance.
 6. Avoids putting the instrument in the area of high temperature, the dust place, moist or the temperature and
the humidity sudden change place;
 7. In this handbook where appears "*" mark, warned the user must
observe, in order to avoid influence curative effect and damage of the
equipment;


Chapter Two - operating instructions
2.1 Main machine and modules
1 main machine    10 batteries lids
2 Windows         11 power sources plugs
3 power switches  12 eyeshade  (eyepatch)
4 power indicator 13 antidazzling adjust switch  
5 work switches   14 pupils distance adjust knob
6 work indicator  15 earphone
7 earphone jacks  16 adapters
8 functions change-over switch
9 speakers outputs hole

* Attention: avoids using other specifications to
substitute the above module.

2.2 Installation prompt
 1. Connects the power source
  A. When use outside power source, the voltage adapter and the main engine
connection, puts through the power source. (In ordinary circumstances,
suggestion use voltage adapter; * This instrument rated voltage is
220V, If is the country rated voltage is 110V in your country, please first insert the transformer
does not have the segment if applicable, then have
the voltage adapter with the segment inserted in its above)
  B. When uses the battery, level installs the battery according to "+" "-" in the instrument battery box;

  2.Establishment switch position: According to the perpose of the use, switch the function
change-over switch to "A" or "B" ;
  "A" position: The single eye mode - treats the young people adjustable
nearsightedness;
                The binocular - eliminate eyestrain, prevents ambliopia and
the nearsightedness, delays the presbyopia;

  "B" position: The single eye mode - treats types of children ambliopia

 3. Adjustment of antidazzling board
  A. Single eye treatment, eye care: adjust the antidazzling board, to block one eye,
adjust to the opposite position when change eyes;
  B. Binocular treatment, eye care: adjust the antidazzling board to
the position of both eyes all sees in the target image in the Windows;

 4. Installment protection blinkers eyeshade(eyepatch)
    Insert the eyepatch in the windows;

 5. The adjustment of the elevation angle
    The user may adjust the elevation angle appropriately according to the environment;

 6. Installs the earphone
    If you do not want to affect other people then you can switch off the speaker by plug the
earphone in the earphone jack, the speaker sound automatic shut-off, adjust on the earphone the volume
switch to properly volume;

 7. The pupil distance adjustment
    Each person' pupil distance is not same, when at the same time both eyes observes in the Windows ,
please adjust the knob untill observe to the image superposition with both eye.

* Attention: The protection blinkers eyeshade material is soft, the
packing and the laying aside please do not extrude, in order
to avoid distortion.


2.3 Power on  procedure
1. Power on machine: Presses the power switch, the power indicator light shines
(green light expression normal work condition, red candle expression
battery capacity is insufficient)

2. Work: Lightly touches the work switch, work indicator to
shine, the instrument starts the treatment procedure which enters the
microcomputer control; Penetrates the Windows to be possible to see
to glitters the change color image. Image will change from the fixed frequency which starts transits gradually
to music frequency shift, it has 25 musics to select the synchronized broadcast immediately. After a treatment
procedure completes automatically stops, the instrument is in awaits orders condition.

3. Termination work:
   A) completes after a treatment procedure, in the Windows the image
glitters stops, if confiscation music has not  been able to play
finished then can continue to play music, when the music is finished it will stop automaticly,
work indicator light extinguish automatically. if you want the
music to terminate ahead of time, simply touches the work switch slightly,working indicator light will extinguish.

   B)Power off in the work midway, lightly touches the work switch;

4. Power off: Presse down the power switch, the power light extinguishes,
tears off the power source adapter and the transformer;

2.4 application methods

1. Method of treatment: put the eye which needs to treat or the health care eye sticks to
eyepatch gently, reduces the outside light as far as possible the disturbance. Penetrates the Windows, please consentrate
on the image inside, Until a working routine automatic conclusion. (In the darkroom
uses, when concentrate in the entire course of treatment, and the
spirit relaxes, may achieve the best treatment result. )

2. Treatment time:
  "A" position: a working routine is 7 minutes, treat both eyes at the same time when uses in the health care,
a working routine is a treatment. When the simple eye use uses in the treatment, two eyes each
becomes a working routine (each eye to use 7 minutes) is a complete treatment.
  "B" position: A working routine is 5 minutes. Only uses in treating
ambliopia, each time treats, must treat the eye should complete
two working routines (each eye to treat the 2x5 minute) continuously,For a treatment.

   Uses for treatment: Every day uses 2-3 time;
   Uses for eye care: Every day uses 1-2 time. Each time treats the time-gap not to be shorter than 4 hours.

* Attention: After several days treatment, the eye can have the slight sour pain, ambliopia patient occasionally
to be able to appear the slight dizziness phenomenon, by now treated the fruitful performance, this was so long as
continued to persist according to the stipulation request treatment, the above phenomenon could vanish gradually.

3. Treatment cycle: 15 days are a treatment cycle. Young people's adjustable nearsightedness treats needs 2-3 treatment
 course; The children ambliopia treatment will need 5-8 treatment course; Regarding consolidated curative effect,
 eliminate eyestrain, delays the presbyopia, should use long term.

* Attention: This machine is uses the human body vision frequency
spectrum light exercise and the eye massage, persisted uses may make
the eye to maintain the health, the long-term use does not have any side effect.


 4. Treatment standard
    A. Effective: After the process treatment, the result of single eye  or corrective vision
(refers weakly regards patient),enhances two line or above two lines(eyechart comparison) upon the original foundation.
    B. Cure: After the process treatment, the result of single eye  or corrective vision, to
achieve above 1.0 or 1.0 for restore normally.
    C. Invalid: After the process treatment, the vision does not have the
enhancement when compare with an eyechart under the smae circumstance.

 5. Attentions
    A. Ambliopia patient suppose after the examination the eyes
and has one's eyes tested and wear the eyeglasses to
treat, and obeys the doctor's advice;
    B. Nearsightedness, presbyopia patient and so on when treatment must take
off the eyeglasses; Mild shortsighted patient in treatment period
should as far as possible not to wear the eyeglasses often, will cause a quicker enhancement and stable;
    C. All different types of patient needs periodic inspection, when
the vision enhanced or the diopter already reduces, should change adapt eyeglasses promptly;
If vision already restored normally, should take off the eyeglasses.
    D. Treatment period should reduce with the eye time, participates in the
outdoors activity, do not watch the TV as far as possible, do not play the computer or TV games;
    E. After uses the vision has the distinct enhancement,make longer the time interval between each treatment course;


Chapter Three - product maintenance

3.1 product maintenance
Windows lens clean: When clean Windows lens please use the lens paper
to turn circle from in to the outside counter clockwise carefully.

* Attention: If the lens are mildewy when either other pollution is
unable clean or the damage, should return to the seller for service, do
not disassemble arbitrarily;
       Protects the blinkers eyeshade cleanly: the eyepatch needs to be take off often, and clean with
the medium detergent;
       Outer covering clean: When clean device cover, must power off first (tear off
transformer and adapter), uses softly cloth to dip the neutral detergent to clean it; Prohibition use
ethyl alcohol or gasoline and so on organic solvent. Careful do not have to splash clean liquid on other
parts of the device or in your eyes.

3.2 Technical specification

 1. Environmental condition:
    Operating temperature: 5.C - 40.C   work humidity: <80%
    Storing  temperature: -10.C-50.C   storing humidity: <80%

 2. Volume and weight
    Volume: Long x width x high =285mm x 195mm x 310mm
    Weight: 1.5kg

 3. Rated power: <2W

 4. Module and fitting:
    Voltage adapter: Input voltage: AC 220V+-10%/50Hz
                     Output voltage: DC 6/300mA


 5. Stereo earphones:adjustable volume, impedance 32 ohms, frequency loud scope 20-2kHz

    Note: If there will be any change in the technical specification of the design and the disposition of this product,
we will not informe you separately;


3.3 trouble shootings

trouble phenomenon               possible trouble reason               trouble shooting method

1. The power light is not on,   The power supply is not normal,       Inspection voltage adapter or battery whether do install correctly
2. The opening machine working is abnormal,  the battery voltage is lower than the normal value, Replaces with the new battery
3. The opening machine working is abnormal,  Operates is improper,      power off to check and power on again
4. Does not have the blue light,     The battery voltage excessively is low,  Replaces with the new battery

-eof